The purpose of surgical trays is to store the instruments used to insert dental implants. The kit can be carried, sterilized and kept in a horizontal position with the lid closed. All the instruments must be cleaned and sterilized before the first use. The surgical kit and instruments are not sterile at the time of delivery.

Cleaning

1] Dismantle all the compound parts.
2] Rinse abundantly with cold or lukewarm water for 2-5 minutes.
3] Leave the instruments for 10 minutes in an ultrasonic cleaner with a neutral pH. enzymatic detergent diluted in water according to the product instructions.
4] Wash the instruments with water for 3 minutes.

Sterilization

The guidelines for sterilisation are listed below.
Exceeding these sterilisation limits may cause deterioration of the plastic components.

type of cycle (value)

temperature (°C/F)

exposure

drying time

Pre-vacuum

132 / 270

3 minutes

30 minutes

Pre-vacuum

134 / 273

18 minutes

30 minutes

Gravity

121 / 250

80 minutes

30 minutes

Prosthetic kit composition

Code

Description

MDS

short screwdriver hex 1.25 mm h 23 mm

MDL

long screwdriver hex 1.25 mm h 29 mm

TRT

removal tool for abutments

HDH20

implant driver for straight MUA

THDDS

short prosthetic screwdriver hex 1.25

THDDL

long prosthetic screwdriver hex 1.25

TWA2

ratchet wrench

MDLA (on request)

angled screwdriver hex 1.25 mm h 29 mm